Analyzer, Electrolyte, 5 Touch, ISE-4000, Na+, K+, Cl & HCO3
- SFRI ISE 4000 use the Van Slyke Manometric method of analysis to determine total CO2, by measuring the pressure of CO2 in samples after reaction with lactic acid.
- TCO2 calibration is very stable thanks to stable reagents and a simple analysis method.
- No expensive and unstable enzymatic reagents are required for this test ensuring more reliable results.
- Excellent correlation between Manometric method and standard enzymatic method ensures results are precise and reproducible.
Description
Optimized Efficiency
Thanks to its automatically performed 2 point calibration every 4 hours, results are accurate with good reproducibility through time (automatic curve compensation is available). 1 point calibration is also ensured with every sample. Optional barcode reader allows for quick and easy identification of samples.
User-friendly Operations
The interactive touch screen, large display and user-friendly interface make for a trouble-free and effortless working experience. All ISE models are equipped with a 24 hour non-stop working mode suitable for emergency sample analysis. The built-in operating system alerts operators when reagent levels are low or a new pack is required. Electrodes and reagent packs snap in and out for easy replacement. Error alerts appear on print-out if values are abnormal.
Automation and Auto Sampling
Automation for high productivity is guaranteed by automatic control software, aspiration, washing and calibration. Results display and print-out automatically. All models can be upgraded to ensure higher automation thanks to the optional auto sampler. Operators can program a series of up to 20 tests at one time all the while being able to run emergency tests if necessary. All results can be transferred to patient files for storage via mono directional LIS. With its programmable tests and disposable cups the auto sampler makes for a hassle-free work process.
Reagent Pack
The ISE SERIES is manufactured with a built-in disposable all-in-one reagents pack for quick and easy replacement of reagents. Each Pack contains 1 L of solution and is automatically identified by RFID. No liquid handling is necessary; no risk of hook up errors can be made. The waste bag is integrated in the Pack eliminating any risk of contamination.
Easy and Simple Maintenance
The advanced design of hardware, fluid tubing system and self-diagnosis software make for easy and simple maintenance and troubleshooting.
Low Operating Costs
The excellent quality of electrodes, high stability of reagents and low consumption rate make the ISE SERIES a very cost efficient analyzer.
Productivity and precision
The ISE renders results in only 60s and can store up to 200 patient files every 24h. It uses only 150 µl of sample volume and measures directly without dilution for faster measuring time. With its low CV, wide linear range, high performance electrodes and unique calibration programs designed to eliminate systematic errors, the ISE models are reliable and accurate.
Technical Specifications
Principles | Ion Selective Electrode direct measurement Manometric method for TCO2 |
Samples | 150 µl: Whole blood, serum, plasma and urine |
Throughput | 60 tests per hour without TCO2 45 tests per hour with TCO2 |
Detection Range | Whole blood, serum and plasma K+ : 0.5 – 15.0 mmol/L Na+ : 20 – 200 mmol/L Cl- : 20 – 200 mmol/L Li+ : 0.2 – 3.0 mmol/L Ca2+ : 0.1 – 5.0 mmol/L pH : 4.0 – 9.0 pH TCO2 : 6.0 – 50.0 mmol/L Urine K+ : 5.0 – 250 mmol/L (pre-diluted) Na+ : 10 – 350 mmol/L Cl- : 10 – 350 mmol/L |
Quality Control | Two levels with auto sampler One level without auto sampler |
Calibration | Automatic: 2 points every 4h or on demand |
Memory Storage | 200 patient files x 5 parameters each: 1000 results every 24h Mono-directional LIS transfer |
Reagents Pack | Shelf life: 12 to 18 months Standard Pack: Cal A, Cal B, Waste Pack for ISE 4000 & 6000: Cal A, Cal B, TCO2 reagent. |
Electrodes | Long life: up to 5 years (6 months for Ca2+) 12 month warranty (6 month warranty for Ca2+) |
Input/output | Barcode interface (optional barcode) RS232 port for mono-directional LIS |
Printout | Fast thermal printer, 57.5 mm wide paper, recording width 48 mm |
Display | 5” LCD touch screen with mono-color display |
Operating Environment | Temperature 15°C – 32°C Humidity < 85% (max humidity) |
Power Requirements | A.C. 110/220 V ±10%; 50 – 60 Hz Consumption: < 60 W |
Dimensions | 390(W) x 445(H) x 340(D) mm |
Weight | Main unit: 10 kg; Auto sampler: 1.5 kg |
For detailed specifications refer to the Product Brochure.
SFRI
ISE 4000 Electrolyte Analyzer (Na+, K+, Cl-, TCO2)
The ISE 4000 is equipped with the TCO2 module; an invaluable function that allows for the analysis of bicarbonates in the blood. This test measures the level of bicarbonates (HCO3 – ) in a sample of blood or urine. It is usually ordered in addition to the standard electrolyte panel to measure the severity of an acid base imbalance in the blood, determine its origin (respiratory or metabolic) and monitor treatment. A carbon dioxide (bicarbonate) test helps find and keep track of conditions that affect blood bicarbonate levels, including many kidney diseases, some lung diseases, and metabolic conditions.